Dew Point

The Dew Point Sensor NP330- G

The Dew Point Sensor NP330-GThe Dew Point Sensor NP330- G  is a 4-20mA transmitter to measure absolute moisture in °C dew point (°C dp) in gases.

Due to a special manufacturing process it has a high media compatibility for use as online unit for the most diverse gases and gas mixtures for control, monitoring and quality assurance in gas treatment. Due to its wide measuring range the sensor is used for monitoring and control of freeze dryers and adsorption dryers.

The use of the NP330-G ensures the quality of air dryers in production processes and saves energy during drying.

 

 

technical data

Measuring range Dewpoint:

-80 °C dp - +20 °C dp

Measuring range Temperature:

-20 °C - +60 °C

Measuring accuracy:

± 2 °C dp

Operating Pressure::

Vakuum bis 30 MPa

Temperature Compensation:

 available

Operating Temperature:

-20° C - +60° C

Medium compatibility:

ther most gases and gas mixtures

Protective sleeve:

Sinter guard filter, stainless steel 80 µm

Operating voltage:

18 – 28 V DC, ca. 30 – 70 mA

Outputs:

Current output 4-20 mA, max. BĂĽrde 1kOhm
RS 232 Data I/O for Computer network
Switch output (optional) 24 V DC, max. 100 mA

Connector:

Circular connector, 7-pins,
Typ Binder 723

Protection:

IP 65

Sample connecting thread:

5/8 – 18 UNF, SW 22 mm

Depht:

50 mm

High of housing:

115 mm

Housing diameter:

60 mm

Length:

176 mm

Mass:

450 g

Recalibration:

Time intervals dependent on the employment an annual examination is recommended.

 

Dew point switch  typ  NP 400

Dew point switch typ NP 400The dew point switch  typ  NP 400 is designed according to the same principle as the sensor above. However, as a low cost unit, it is especially intended to monitor  moisture in gases.

This is achieved by an electronic unit integrated in the sensor head and contains two adjustable outputs. A preset fixed switch point is triggered when the moisture is exceeded or falls short of it.

 

technical data

Measured variable:

Absolutfeuchtegehalt in Gasen

Measuring range:

Typ 1: -46 ... -14 °C TP
Typ 2: -15 ... +15 °C TP
Typ 3: -76 ... -46°C TP

Accuracy:

± 3 °C TP

Resolution:

< 0,5 °C Tp

2 switch points:

variable

switch hysteresis:

variable in the range 1 ... 6 °C TP

Time delay:

variable in the range 1 ... 240 sec

Circuit logic:

active low (s the current dampness more largely than the adjusted threshold is the appropriate exit up Low-Pegel)

Operating pressure:

Vakuum to 30 MPa

Operatingtemperature:

0 ... +60 °C

Operating voltage:

24V DC +/- 15% , ca. 15 mA internal requirement

Protection:

IP 65

Output:

operating voltage switches at exit,
max. 500 mA

Connector:

4 pin. Connector IEC 947-5-2 DIN/VDE 0660

Connection allocation:

Pin 1: output 1 (green LED)
Pin 2: power supply, +24V
Pin 3: mass, 0V
Pin 4: output 2 (yellow LED)

Connecting Cable:

Pin 1 - brown
Pin 2 - white
Pin 3 - blue
Pin 4 - black

Sample connecting thread:

1/2 ", SW 27

Depht:

ca. 55 mm

Lenght:

ca. 138 mm

Mass:

ca. 105 g

Rekalibration:

Time intervals dependent on the employment an annual examination is recommended.

 

 

For further information and ordering, please click on the link Contact on this page and request an offer.

You can also reach us over the phone 0049 35952 / 42 94 65.